Skip to main content
Journal of Foot and Ankle Research logoLink to Journal of Foot and Ankle Research
. 2015 Sep 22;8(Suppl 2):P12. doi: 10.1186/1757-1146-8-S2-P12

Reliability of a smartphone goniometer app compared with traditional goniometer for measuring passive motion at the first metatarsophalangeal joint

Georgie Hales 1, Richard Keating 1, Nicola Bear 1, Katie Warren 1, Simon Otter 1,
PMCID: PMC4595090

Background

Adequate sagittal plane motion of the first metatarsalphalangeal joint (1st MTPJ) is important during normal gait and goniometric measurement is commonly used as a diagnostic and outcome assessment tool for 1st MTPJ pathology. We aimed to determine the intra and inter-rater reliability of a universal plastic goniometer and a smartphone application when measuring dorsiflexion at the 1st MTPJ.

Method

A double-blind repeated measures design was used (three raters, one observer) to measure 1st MTPJ dorsiflexion in both feet of 32 healthy volunteers (64 data sets) using a pre-defined measurement protocol. Reliability of the smartphone app (Dr. Goniometer v1.8 installed on an iphone 4s) and a standard universal goniometer was assessed using Interclass correlation coefficients (ICCs) with 95% confidence intervals.

Results

Both instruments demonstrated excellent intra-rater reliability (ICC> 0.75), with moderate to good inter-rater reliability:

Universal Goniometer

Inter-rater reliability ICC 0.693 (95% CI 0.580 – 0.788)

Smartphone app

Inter-rater reliability ICC 0.708 (95% CI 0.597 - 0.799)

Conclusions

Moderate to high intra and inter-rater reliability of passive 1st MTPJ motion can be achieved with traditional and smartphone-based goniometric measurement. Smartphone applications may provide a slightly higher reliability, but devices should not be used inter-changeably as significant variations in measurement between devices may occur.


Articles from Journal of Foot and Ankle Research are provided here courtesy of Wiley

RESOURCES