Fig. 8.

Fig. 8

Comparison of T1ρ maps errors obtained from reconstructions of combined (T1ρ + T2) dataset and the T1ρ only dataset: The T1ρ maps errors at different accelerations for all the schemes on the combined dataset (solid lines) and only T1ρ dataset (dotted lines) are shown. The plot on left shows comparisons for the datasets without any motion and the plot on the right shows comparisons for datasets with motion. We observe in both cases that BCS performs better than CS and kt-PCA schemes. In other words combining the datasets improves the reconstructions.