TABLE III.

Comparison results of heterogeneous polyenergetic DgN (pDgNhetero) values for displacement of the glandular distributions.

Clinically relevant mammography spectra
Phantom (VGF) Target /filtera kV HVL (mm Al) Glandular distribution displacement pDgNhetero (mGy/mGy) Change (%) (relative to centered)
Medium (12.6%) Mo–Mo 31 0.4 Centered 0.169
Superior 0.232 37.3
Inferior 0.124 −26.6
Lateral 0.173 2.4
W–Rh 31 0.5 Centered 0.263
Superior 0.338 28.5
Inferior 0.205 −22.1
Lateral 0.267 1.5
a

Mo–Mo (0.03 mm), W–Rh (0.05 mm).