Table 1.

Edge-to-edge distances between resolution cavities in CBCT scans and estimated width and length of the imaging phantom. Results are presented as deviation from the true values, 3 mm between air cavities, 14 mm between corner air cavities, 25.4 mm phantom diameter and 80 mm phantom length.

Dates of
QA check
Deviation in distances between
air cavities in transverse/sagittal
slices (mm)
Deviation in
distances between
corner air cavities
(mm)
Deviation in
phantom
representation
(mm)

1mm→
2mm
2mm→
3mm
3mm→
4mm
Horizontal Vertica
l
Diameter Lengt
h

January 6 0/+0.3 +0.6/+0.6 +0.3/+0.3 −0.5 0 −0.1 −0.2
January 7 +0.3/0 0/+0.3 +0.3/+0.3 −0.8 −0.2 −0.4 −0.2
January 9 0/0 +0.3/+0.3 +0.3/+0.3 −0.8 +0.3 +1.1 −0.5
January 10 0/+0.3 +0.6/+0.3 +0.3/+0.6 −0.8 +0.3 +0.2 −0.2
January 16 +0.3/+0.6 +0.3/+0.6 +0.6/+0.3 −0.8 +0.3 −0.1 −1.1
Isocenter re-alignment and system re-calibration on January 23
January 29 0/0 +0.3/+0.3 +0.3/+0.3 −0.2 +0.3 +1.0 +0.3
February 7 0/+0.3 +0.3/+0.6 +0.6/+0.3 −0.2 +0.3 +1.0 0
February 14 +0.6/+0.3 +0.3/+0.6 +0.3/+0.3 −0.5 0 +1.0 +0.6
March 3 0/0 +0.3/+0.3 +0.3/+0.3 −0.2 +0.3 +0.2 0
March 18 0/+0.3 +0.3/+0.3 +0.3/+0.6 −0.2 0 +0.7 +0.3
April 25 0/+0.3 +0.3/+0.3 +0.3/+0.3 −0.2 0 +0.2 +0.3
April 3 0/+0.3 +0.3/+0.3 0/+0.3 −0.2 −0.2 −0.1 0
May 5 0/+0.3 +0.3/+0.3 +0.3/+0.3 −0.2 +0.3 +0.5 0
May 15 0/0 0/+0.3 0/+0.3 −0.2 +0.3 −0.1 +0.3
May 29 0/+0.3 +0.3/+0.3 +0.3/+0.3 −0.5 +0.3 +0.2 +0.3