Published in final edited form as: Health Phys. 2015 Nov;109(3 0 3):S190–S199. doi: 10.1097/HP.0000000000000323

Table 2.

Tests included in the proposed quality management program and recommended tolerances for long-term consistency measures.

Performance Tests Tolerance
Output consistency Measure dose-rate the at isocenter using
an appropriately calibrated ion chamber
± 1%
Image resolution
consistency /
Object representation
Use a CBCT scan of the imaging
phantom to derive diameters of all
resolution air cavities by vertical and
horizontal line profiles. Derive
distances between cavities from the
same line profiles. Check the length and
diameter of the imaging phantom in a
sagittal slice.
± 0.2 mm for
resolution
± 0.5 mm for
distances
± 1.0 mm for object
size representation
Accuracy of image-
guided target
localization
Locate a well-defined target using the
high-CT density BB phantom and then
verify that its location identified on the
CBCT coincides with the radiation
isocenter using the 5 × 5 mm2 collimator
at gantry angles 0° and 90°.
± 0.5 mm
Dose calculation
consistency
Calculate a four-field 10 × 10 mm2
treatment plan for 10 Gy on a
reproducible isocenter in the imaging
phantom and record the treatment
times.
± 2 s (± 3%) per
treatment field