Gradient Pump (Dionex GP50)
Mobile Phase Composition: 88% Deionized Water & 12% 250 mM Sodium Hydroxide
Flow Rate: 0.4 mL/minute
High Pressure Limit: 2,000 psi
-
Isocratic Pump for Rezex RCM Monosaccharide Columns (Bio-Rad 1350T)
Mobile Phase Composition: 100% Deionized Water
Flow Rate: 0.6 mL/minute
High Pressure Limit: 256 kg/cm2
-
Isocratic Pump for Dionex PA1 Column Cleanup (Bio-Rad 1350T)
Mobile Phase Composition: 1.0 M Sodium Hydroxide
Flow Rate: 0.4 mL/minute
High Pressure Limit: 256 kg/cm2
-
Column Heater
Temperature: 65°C
-
Autosampler (Dionex AS50)
Injection Volume: 20 μL
Run Time: 30 minutes
-
Electrochemical Detector (Dionex ED50)
Analog Range: 1 μC
-
Waveform:
0.00 sec, E = +0.05 V
0.20 sec, E = +0.05 V
0.40 sec, E = +0.05 V
0.41 sec, E = +0.75 V
0.60 sec, E = +0.75 V
0.61 sec, E = −0.15 V
1.00 sec, E = −0.15 V
Integration Period: 0.20 – 0.40 seconds
gold electrode
|