Table 4.

Configuration of ARIA III.

Detector Long-pass filter Band-pass filter Fluorochrome
Blue laser (488 nm)
A 655LP 695/40 PerCP-Cy5.5
B 502LP 530/30 FITC
C 488/10 SSC
Yellow/green laser (561 nm)
A 735LP 780/60 PE-Cy7
D 600LP 610/20 PE Texas Red
E 582/15 PE
Red laser (640 nm)
A 735LP 780/60 APC-Cy7
B 690LP 730/45 Alexa Fluor 700
C 660/20 APC
Violet laser (407 nm)
A 600LP 660/20 Qdot 655
B 502LP 510/50 BV510
C 450/40 BV421