Published in final edited form as: Diabetes Care. 2008 Mar 13;31(6):1160–1164. doi: 10.2337/dc07-2401

Table 1.

Numerical accuracy

Guardian DexCom Navigator Glucoday
Point accuracy: euglycemia (blood glucose = 3.9–10 mmol/l)
 MAD (mmol/l) 0.91 1.24 0.89 0.87
 MARD 15.2 21.2 15.3 15.6
 MedAD (mmol/l) 0.82 1.06 0.85 0.60
 MedARD 13.3 18.4 11.8 10.7
 ISO: % readings within 20% from reference when reference >4.2 mmol/l 73.2 52.2 72.2 76.9
 MAD (mmol/l) 0.55 0.73 0.36 0.47
 MARD 16.1 21.5 10.3 17.5
 MedAD (mmol/l) 0.42 0.64 0.24 0.40
 MedARD 13.8 22.5 7.4 15.6
 ISO: % readings within 0.8 mmol/l from reference when reference ≤4.2 mmol/l 76.5 52.9 79.4 83.0
Rate accuracy: descent into hypoglycemia Absolute R-deviation (mmol · l−1 · h−1) 2.9 2.4 2.2 5.8
Rate accuracy: ascent from hypoglycemia Absolute R-deviation (mmol · l−1 · h−1) 3.0 3.3 3.3 9.3

Data are percent unless otherwise indicated.