Fig. 3.
Bland-Altman plot of the outer retinal thickness, the RPE_OScomplex, for all measurement points with the Cirrus OCT and the Heidelberg OCT. The y-axis shows the difference in thickness (μm) between measurements obtained using the Heidelberg OCT and Cirrus OCT, and the x-axis shows the mean sum of the two instruments. The mean difference of 3.32 μm is shown with a stippled line, and the mean ± 2SD is shown with solid lines