. 2015 Jun 12;253(11):1915–1921. doi: 10.1007/s00417-015-3075-2

Table 1.

Retinal thickness as measured with third-party software on OCT scans from the Cirrus HD and Heidelberg Spectralis systems. Results are reported using means and SDs for the foveal center (top), the central 1-mm line crossing the fovea (middle), and all points of the 6-mm line (bottom). The difference between the instruments is given in μm, and the percentage and the p value are also for comparison of the instruments

Retinal Thickness Cirrus HD Heidelberg Spectralis Difference % Difference p value
μm μm μm
Foveal center, mean 225.92 228.70 2.78 1.22 0.0555
SD 17.02 18.42 8.56
1-mm line, mean 280.37 282.15 1.78 0.63 0.0414
SD 17.90 18.40 5.13
All points, mean 301.10 302.95 1.85 0.62 0.0052
SD 39.07 38.57 7.50