TABLE 2.
Simulation Results Demonstrating the Effect of Probe Size on the pO2 Readout
Method of measurement | Skewed distribution
|
Normal distribution
|
||||
---|---|---|---|---|---|---|
Mean (mmHg) | Median (mmHg) | Skewness | Mean (mmHg) | Median (mmHg) | Skewness | |
Actual (input) | 1.50 | 0.91 | 2.31 | 7.59 | 7.59 | 0 |
Microelectrode | 1.53 | 0.89 | 2.1 | 7.61 | 7.61 | 0.12 |
OxyLite | 1.47 | 1.39 | 1.22 | 7.60 | 7.60 | 0.14 |
EPR | 1.51 | 1.49 | 0.37 | 7.60 | 7.58 | 0.09 |