. 2015 Oct 16;53(11):3458–3465. doi: 10.1128/JCM.01806-15

TABLE 1.

Concordance, Pearson correlation coefficients, and numbers of samples with viral load differences of >0.5 log copies/ml between assays

Assays compared and test 1 viral loads No. of samples with test 2 viral loads of:
Total no. Concordant samples (%) No. of quantified samples Pearson r % (no.) of samples with differences of:
≤1.59 log copies/ml >1.60 log copies/ml >0.5 log copies/ml <0.5 log copies/ml
NucliSens vs RealTime 89.74 162 0.91 10.49 (17) 20.99 (34)
    ≤1.59 log copies/ml 83 16 99
    >1.60 log copies/ml 12 162 174
NucliSens vs Xpert 84.96 146 0.90 4.79 (7) 26.71 (39)
    ≤1.59 log copies/ml 80 18 98
    >1.60 log copies/ml 22 146 168
NucliSens vs Aptima 83.89 163 0.89 5.52 (9) 38.65 (63)
    ≤1.59 log copies/ml 87 34 121
    >1.60 log copies/ml 14 163 177
RealTime vs Xpert 89.80 82 0.98 0 (0) 10.98 (9)
    ≤1.59 log copies/ml 50 3 53
    >1.60 log copies/ml 12 82 94
RealTime vs Aptima 89.78 82 0.98 3.66 (3) 20.73 (17)
    ≤1.59 log copies/ml 41 11 52
    >1.60 log copies/ml 3 82 85
Xpert vs Aptima 91.37 116 0.97 3.45 (4) 6.90 (8)
    ≤1.59 log copies/ml 64 4 68
    >1.60 log copies/ml 13 116 129