Table 1.

Comparison of apparent diffusion coefficient (ADC) values.

b-value seta Mean error relative
to ADCIVIMb (%)
p-valuec Intra-scan
CVd (%)
0–1000 18.6 <0.001 18.6
50–1000 13.0 <0.001 19.1
100–1000 9.4 <0.001 19.8
250–1000 2.2 0.564 22.4e
500–1000 −7.8 <0.001 29.4
250, 500 and 1000 −0.13 0.402 23.1e
250, 650 and 1000 1.6 0.805 22.6e
250, 800 and 1000 3.5 0.113 21.9e
250 and 800 12.5 <0.001 24.9e
250 and 1000 1.3 0.994 22.7e
0–1000 (IVIM) - - 24.1
a

The b-value (in µs/µm2) sets were used for the monoexponential model except the last row which used IVIM model;

b

ADCIVIM represents the IVIM model ADC;

c

Two-tailed paired t-test with ADCIVIM mean values of the active tumour volumes of the scans;

d

CV, coefficient of variation;

e

Not significantly different intra-scan CV of ADC values of the active tumour volumes compared to that of IVIM (two-tailed paired t-test).