. 2015 Sep 30;68(5):455–461. doi: 10.4097/kjae.2015.68.5.455

Table 2. Comparative Data and Complications on Airway Devices.

i-gel (n = 47) SLMA (n = 46) P
Size (3/4/5) 5/42/0 7/35/4 0.08
Attempt No. (1/2/3) 46/1/0 45/0/1 0.37
Insertion time (sec) 15.7 ± 3.7 15.7 ± 3.1 0.99
Fiberoptic view (1/2/3/4) 0/3/16/28 1/9/10/26 0.14
Gastric tube insertion time (sec) 20.4 ± 3.9 16.7 ± 1.6 < 0.001
Audible air leak 5 8 0.39
Manipulation frequency (0/1/2) 43/4/0 39/6/1 0.45
Complications
 Blood on device 3 2 0.51
 Sore throat 4 1 0.19
 Others 2 2 1.00

Data are presented as mean ± standard deviation, or numbers. Fiberoptic view: 1, vocal cords not seen; 2, vocal cords plus anterior epiglottis; 3, vocal cords plus posterior epiglottis; 4, vocal cords only. SLMA: Supreme laryngeal mask airway.