Skip to main content
. Author manuscript; available in PMC: 2016 Oct 21.
Published in final edited form as: Phys Med Biol. 2015 Sep 25;60(20):7927–7939. doi: 10.1088/0031-9155/60/20/7927

Table 1.

Average (Avg) and standard deviation (Std) of the measured error on the output factor determined for 7 repeated measurements at different field sizes and depths under warm conditions. Output factors were normalized to the value of the 10 × 10 cm2 field size at 0.5-cm depth.

Field size
5 × 5 cm2 10 × 10 cm2 15 × 15 cm2
Depth Avg Std Avg Std Avg Std
0.5 cm 0.09% 0.49% 0% 0.65% −1.17% 0.85%
10 cm 1.23% 0.31% −0.21% 0.29% −1.88% 0.34%