Table 1.

Spatial resolution of current CBCT scanners as published in recent literature

Publication Type of phantom Visual resolution (lp mm−1)
MTF at 10% modulation (cycles per millimetre)
Voxel size (range in millimetre)
CBCT device(s)
Min Max Min Max Min Max
Suomalainen et al42 Teflon® edge vs Perspex® background na na 0.10 0.80 0.120 0.360 3D Accuitomo
PromaxH 3D scanner
Scanora® 3D scanner
Xu et al43 Wire phantom 1.20 1.6 1.80 2.30 0.200 0.500 CS 9300
Ozaki et al44 Thin tungsten wire 1.70 2.8 1.80 2.60 0.080 0.160 3D Accuitomo FineCube v. 12
Bamba et al45 SEDENTEXCT, thin stainless steel vs air 1.00 2.8 0.60 1.25 0.080 0.125 CS 9300
3D Accuitomo 80
Veraviewepocs
Ballrick et al46 Line-pair phantom [C phantom (Phantom Laboratory, Salem, NY)] 0.60 0.8 na na 0.200 0.400 i-CAT model 9140-0035-000C
Knörgen et al47 Edge PMMA vs air 0.70 1.7 na na 0.125 0.250 3D Accuitomo 170
Pauwels et al48 SEDENTEXCT, line-pair insert (aluminium vs polymer) na <3.0 na na 0.076 0.400 3D Accuitomo 170
3D Accuitomo (image intensifier)
Veraviewepocs
Galileos Comfort
i-CAT Next Generation
CS 9000 3D
CS 9500
NewTom VGi
Pax-Uni3D
Picasso Trio
ProMax® 3D
Scanora 3D scanner
SkyView®
Steiding et al9 Edge aluminium sphere vs foam 0.83 1.0 0.88 0.89 0.300 0.300 i-CAT Next Generation
Watanabe et al49 0.100-mm tungsten wire; MTF, 0.100-mm tungsten wire; visual, aluminium vs epoxy resin 1.90 2.1 1.75 2.64 0.125 0.125 3D Accuitomo (image intensifier)
Watanabe et al50 MTF, 0.100-mm tungsten wire; visual, aluminium vs epoxy resin 1.90 2.1 1.89 2.65 0.125 0.125 3D Accuitomo (image intensifier)

max, maximum; min, minimum; MTF, modulation transfer function; na, not available; PMMA, polymethylmethacrylate.

3D Accuitomo, Veraviewepocs and FineCube v. 12 obtained from J. Morita MFG Corp., Kyoto, Japan; PromaxH 3D scanner obtained from Planmeca Oy, Helsinki, Finland; Scanora® 3D scanner obtained from Soredex, Tuusula, Finland; CS 9300, CS 9000 3D and CS 9500 obtained from Carestream Health, Rochester, NY; i-CAT model 9140-0035-000C obtained from Imaging Sciences International, Hatfield, PA; Galileos Comfort obtained from Sirona Dental Systems, Bensheim, Germany; NewTom VGi obtained from Quantitative Radiology (QR Srl), Verona, Italy; Pax-Uni3D and Picasso Trio obtained from Value Added Technologies (VATECH), Yongin, South Korea; SkyView obtained from Cefla Dental Group, Imola, Italy.