Skip to main content
. 2015 Oct 12;2015:356401. doi: 10.1155/2015/356401

Table 3.

Optimization of LAMP assay for Salmonella detection by targeting the stn gene.

Conditions The tested range Difference Good yield Selected conditiona
Temperature 60–65°C 1°C 64–66°C 65°C
Time 0–60 min 5 min 15–60 min 25 min
Loop primers 0.5–1.0 μM 0.1 μM 0.8–1.0 μM 0.8 μM
Outer primers 0.1–0.5 μM 0.1 μM 0.2–0.5 μM 0.2 μM
MgSO4 4–8 mM 0.5 mM 6–7.5 mM 7 mM
dNTPs 0.8–2.4 mM 0.2 mM 1.2–1.8 mM 1.6 mM
Betaine 0.5–1.5 M 0.1 M 0.6–1.5 M 1.0 M
Bst DNA polymerase 4–12 units 1 unit 8–12 units 8 units

aThe final concentrations of reagents or temperature and times used in LAMP analysis.

The concentration of FIP and BIP primers are fixed at 1.6 μM.