TABLE I.

Targeting Errors During Needle Insertion [4]

Factor Target error for 120 mm insertion
Needle deflection 6.5±3.5 mm
Imaging artifact of needle 1.6±0.4 mm
Template misalignment 1.5±0.3 mm