Table 2.
Criterium | Minimum requirement | Additional recommendation by the DGK |
---|---|---|
12-lead ECG | Permanent availability | |
Blood pressure measurement | At each bed | Non-invasive blood pressure monitoring in the waiting room, facilities for implementing invasive monitoring |
TTE | Available 24/7, response time <30 min | Dedicated CPU machine |
Rhythm monitoring | At each bed | |
Resuscitation | Dedicated facilities, including defibrillator | |
Transportation with ECG monitoring | Permanently available (if necessary with equipment from the intensive care unit) | CPU-dedicated devices |
Transport ventilator | Permanently available (if necessary with equipment from the intensive care unit) | CPU-dedicated devices |
Laboratory diagnostics | 24-h availability; turn-around time 45–60 min | POCT, turn-around time <20 min |
Blood gas analysis | Available; turn-around time <15 min | Integration in the CPU |
External pacemaker | Permanently available (if necessary with equipment from the intensive care unit) | CPU-dedicated devices |
Exercise stress test, CT | Available within three business days; an appointment must be given upon discharge | Cooperation with external walk-in clinics |
TTE transthoracic echocardiography, POCT Point-of-Care Testing, CT computed tomography