Skip to main content
. 2014 Oct 30;6(5):1347–1355. doi: 10.4161/mabs.29938

Table 2.

Minimum concentration of PS-20 required to prevent the production of medium-sized aggregates during the nebulization of an antibody solution (10 mg/ml in PBS pH 7.2) with three nebulizers

Nebulizer Minimum PS-20 concentration (% w/v)
Omron 10−5
Aerogen Solo 10−4
PARI eFlow 10−3