Table III.

Desiderable Bias for parameters normally distributed.

Parameters Desirable bias Sample B Sample C Sample D Sample E Sample F
p vs sample A p vs sample A p vs sample A p vs sample A p vs sample A
% bias % bias % bias % bias % bias
(95% CLA) (95% CLA) (95% CLA) (95% CLA) (95% CLA)
WBC (109/L) ±5.6% p=0.0535 p=0.2108 p=0.1358 p=0.0534 p=0.0844

RBC (109/L) ±1.7% p=0.1607 p=0.0031 p=0.003 p=0.0025 p=0.0013
−1.5 −1.4 −1 −3.6
(1.6; −4.5) (0.9; −3.7) (1.1; −3.1) (3.1; −10.3)

RDW ± 1.7% p=0.003 p=0,6993 p<0.0001 p<0.0001 p=0,0524
−1.14 −2.8 −1.5
(0.73; −3.01) (0.2; −5.3) (0.22; −3.29)

MCHC (g/dL) ±0.4% p=0.6916 p=0.2122 p=0.1649 p=0.5391 p=0.0051
−2.1
(2.8; −7)

MCV (fL) ±1.2% p=0.0003 p<0,0001 p=0.4929 p=0.0002 p<0.0001
1.05 −2.2 1.6 3.4
(2.68; −0.58) (5.1; −0.7) (4; 0.8) (6.1;0.7)

PLT-I (109/L) ±5.9% p=0.6232 p=0.0070 p=0.0057 p=0.8615 p=0.0020
5.3 8.6 7.8
(17; −6.4) (22.3; −5.1) (24.9; −9.4)

PLT-F (109/L) ±5.9% p=0.2557 p=0.0146 p=0.0043 p=0.1931 p=0.0354
5.3 4 2.1
14.4; −3.8) (14; −6.1) (11.7; −7.5)

MPV (fL) ±2.3% p=0.0214 p=0.8600 p=0.0748 p=0.0217 p=0.1573
−1.7 −1.7
(3.5; −6.9) (3.6; −7)

Parameters Desirable
bias
Sample G Sample H Sample I Sample J
p vs sample A p vs sample A p vs sample A p vs sample A
% bias % bias % bias % bias
(95% CLA) (95% CLA) (95% CLA) (95% CLA)

WBC (109/L) ±5.6% p=0.1957 p=0.1934 p=0.1771 p<0.0001
10.1
(21; −0.7)

RBC (109/L) ±1.7% p=0.0002 p=0.6805 p=0.09 p=0.5104
−2.7
(1.5; −6.9)

RDW ±1.7% p<0,0001 p<0.0001 p=0,0005 p<0,0001
−4 −3.8 2.1 −12.8
(−0.4; −7.7) (1; −8.6) (5.8; −1.5) (2.7; −23)

MCHC (g/dL) ±0.4% p=0.0347 p=0.0088 p=0.0001 p<0.0001
1.8 3.1 −3.8 17.5
(8; −4.3) (11; −4.8) (1.9; −9.4) (21.9;13)

MCV (fL) ±1.2% p=0.0690 p=0.0004 p<0.0001 p<0.0001
−3.2 4 −17.9
(2.4; −8.8) (7;1) (−13.9; −21.9)

PLT-I (109/L) ±5.9% p=0.0036 p=0.0023 p=0.0001 p=0.0014
8.4 4.8 6.9 21.9
(22.4; −5.6) (31.5; −22) (24.1; −10.3) (58.4; −14.6)

PLT-F (109/L) ±5.9% p=0.0037 p=0.0022 p=0.0029 p=0.0012
5.8 8.7 3.7 20.7
(16.4; −4.9) (25.1; −7.8) (15.6; −8.3) (51.3; −10)

MPV (fL) ±2.3% p=0.4375 p=0.0128 p=0.0009 p=0.3022
−4.5 −6
(7.3; −16.4) (4.7; −16.7)

Mean % bias and 95% confidence limit of agreement (95% CLA) of the samples that had a significant p versus sample A.

CLA: confidence limit of agreement; WBC: white blood cells; RBC: red blood cells; RDW: red blood cell distribution width; MCHC: mean corpuscular haemoglobin concentration; MCV: mean corpuscular volume; PLT-I: platelet count using the impedance technique; PLT-F: platelet count using the fluorescent technique; MPV: mean platelet volume.