The results using highly crosslinked polyethylene: a followup interval of 10 years in patients younger than 50 years and older than 50 years

Study Number of patients/number of hips Bearing type Mean age at index surgery (years) Length of followup (years) Number revised Wear (mm/year) Complications
Patients 50 years of age or younger
 Kim et al.* [25] 57/67 Alumina HXLPE 28 (30 and younger) 11 0 0.031 ± 0.004 (mean ± SD) (includes bedding-in period) 1 dislocation
 Babovic and Trousdale [1] 50/54 CoCr HXLPE 39 (50 and younger) Minimum 10 1 0.02 ± 0.0047 (mean ± SD) (did not state if includes bedding-in period) 1 dislocation
 Current study 95/105 CoCr, oxidized zirconium, and ceramic HXLPE 42 (50 and younger) 12 (range, 9–14) 1 0.022 (95% CI, 0.015–0.030) (excludes bedding-in) 3 intraoperative fracture 1 DVT
 Kim et al.* [24] 100 pts C on C, ceramic on HXLPE 44 (50 and younger) 12 (minimum 11) 1 0.031 ± 0.004 (mean ± SD) (includes bedding-in period) 5% squeaking or clicking hips
 Kim et al.* [22] 76/79 Alumina HXLPE 46 9 (range, 7–9) 0 0.05 ± 0.02 (mean ± SD) (includes bedding-in period) 1 dislocation
Patients older than 50 years of age
 Ranawat et al. [37] 91/112 CoCr HXLPE 52 (range, 21–65) 6 (range, 5–7.7) 0 0.043 (± 28 SD) NR
 Johanson et al. [21]§ 51/52 CoCr and HXLPE, CoCr and conventional PE (Durasul; Zimmer) 55 10 1 each group RSA 5/61 stem (8%) revision
 Bedard et al.* [2] 139/150 CoCr HXLPE 56 (range, 25–91) Minimum 10 0 0.05 (includes bedding-in period) 0
 Bragdon et al. [4] 159/174 CoCr HXLPE 60 7–13 NR Minimum 7-year followup: 0.018 ± 0.079 (mean ± SD)
Minimum 10-year followup: 0.01 ± 0.0562 (mean ± SD) (excludes 1-year bedding-in period)
0
 Engh et al.* [12] 116/116 CoCr versus HXLPE or conventional polyethylene 63 10.0 ± 1.8 None for wear in the HXLPE group, 2 for dislocation 0.04 ± 0.06 (mean ± SD) (includes bedding-in period) 2 dislocations
 Garcia-Rey et al.* [14] 45/45 CoCr HXLPE 67 Minimum 10 0 0.02 ± 0.016 (mean ± SD) (includes bedding-in period) NR
 Glyn-Jones et al. [17] 39/39 CoCr HXLPE 68 (range, 52–76) Minimum 10 0 0.033 NR

Methods of crosslinking: * 5 Mrad of gamma irradiation and heated; sterilized with gas plasma (Marathon; DePuy, Warsaw, IN, USA); 10 Mrad of gamma beam irradiation and heat-treated; sterilized with gas plasma (Longevity; Zimmer, Warsaw, IN, USA); 7.5 Mrad irradiation and heat-treated; sterilized to 3 Mrad in nitrogen (Crossfire; Stryker, Kalamazoo, MI, USA); §9.5 Mrad of electron-beam irradiation, melted and annealed; sterilized in ethylene oxide (Durasul; Zimmer, Warsaw, IN, USA); HXLPE = highly crosslinked polyethylene; CoCr = cobalt-chromium; C on C = ceramic on ceramic; PE = polyethylene; NR = not reported; CI = confidence interval; RSA = radiostereometric analysis; DVT = deep vein thrombosis.