Skip to main content
. 2015 Oct 30;16:178. doi: 10.1186/s12882-015-0167-5

Table 1.

Formulas used to evaluate fluxes

Qi = Qb x (1-Ht), Qo = Qi
Mi = Qi x Ci
Mo = Qo x Co
Muf = Quf x Cuf
Mtr = Mi - Mo
C = Ci x Qi/(Qi + RF)
SC = 2 × Cuf/(C + Co)

Ci Concentration in inlet plasma before addition of replacement fluid (ng/mL)

C o Concentration in outlet plasma (ng/mL)

C uf Concentration in ultrafiltrate (ng/mL)

C Concentration in inlet plasma after addition of replacement fluid (ng/mL)

Q b Inlet blood flow rate (mL/min)

Q i Inlet plasma flow rate (mL/min)

Q o Outlet plasma flow rate (mL/min)

Q uf Ultrafiltration flow rate (mL/min)

M i Mass inlet rate (ng/min)

M o Mass outlet rate (ng/min)

M uf Mass ultrafiltration rate (ng/min)

M tr Total mass removal rate (ng/min)

RF Replacement fluid flow rate (mL/min)

SC Sieving coefficient