Table 2.
Reported and additional data from radiofrequency-induced heating tests conducted according to the American Society for Testing and Materials F2182-02a method
| Device | Maximum temperature increase (°C) |
|---|---|
| Phantom contents | +0.2 |
| Keeper | +0.3 |
| Coping A | +0.6 |
| Coping B | +0.6 |
The result for Coping A was reported previously.20
The MR system used was an Achieva 3.0 T Nova Dual (Philips, Tokyo, Japan).