Table 3.
Temperature elevation of the devices at each measurement point
| Device | Maximum temperature increase (°C) |
|---|---|
| Keeper | |
| Point 1 | +1.21 |
| Without the prosthesis | +1.13 |
| Coping A | |
| Point 2 | +1.42 |
| Point 3 | +1.36 |
| Point 4 | +1.36 |
| Without the prosthesis | +1.13 |
| Coping B | |
| Point 5 | +1.30 |
| Point 6 | +1.24 |
| Point 7 | +1.19 |
| Without the prosthesis | +1.13 |
Point 1 is located at the side of the keeper; Points 2 and 5, the upper surfaces of the copings; Points 3 and 6, marginal gingival tissue adjacent to the coping; and Points 4 and 7, the apices of the coping posts.