Skip to main content
. 2015 Jul 9;44(8):20150065. doi: 10.1259/dmfr.20150065

Table 1.

Measurements of doses received by the critical organs in phosphor plate (PSP) system (VistaScan® system; Dürr Dental GmbH, Bissingen, Germany) (in mGy)

 
Tongue
Thyroid gland
Right parotid gland
Left parotid gland
Right crystalline lens
Left crystalline lens
Right maxillary sinus
Left maxillary sinus
Digital intraoral system Mean ± SD Mean ± SD Mean ± SD Mean ± SD Mean ± SD Mean ± SD Mean ± SD Mean ± SD
PSP (mGy) 2.39 ± 0.56a 3.38 ± 0.81a 0.19 ± 0.04a 0.14 ± 0.01a 0.28 ± 0.06a 0.30 ± 0.09a 2.94 ± 0.69a 2.95 ± 0.55a
PSP + lead foil (mGy) 0.82 ± 0.21b 2.36 ± 0.76b 0.09 ± 0.02b 0.06 ± 0.01b 0.12 ± 0.03b 0.16 ± 0.05b 2.43 ± 0.58b 2.45 ± 0.59b

SD, standard deviation.

Different letters within the same column represent statistically significant differences p < 0.05.