Table 1.
Parameters | Group A Consecutive mode RFA (n = 20) |
Group B Switching monopolar RFA (n = 20) |
Group C Switching bipolar RFA (n = 20) |
p-valuea |
p-value A vs B |
p-value A vs C |
p-value B vs C |
---|---|---|---|---|---|---|---|
Technical parameters | |||||||
Total delivered energy (kJ) | 120.9 ± 24.5 | 114.2 ± 18.3 | 76.8 ± 14.3 | <0.001 | 0.728 | <0.001 | <0.001 |
Delivered energy per minute | 5.3 ± 1.1 | 6.5 ± 1.3 | 9.46 ± 1.4 | <0.001 | <0.001 | 0.006 | <0.001 |
Average Watt | 84 ± 17 | 158.7 ± 25.4 | 106.7 ± 19.9 | <0.001 | <0.001 | 0.006 | <0.001 |
Impedance (W) | 57.9 ± 7.2 | 62.7 ± 12.7 | 61.3 ± 11.9 | 0.394 | 1.000 | 0.550 | 1.000 |
Qualitative analysis of coagulation necrosis | |||||||
Confluent necrosis | 30% (6/20) | 85% (17/20) | 90% (18/20) | 0.0002b |
|||
Round shape | 0 | 3 | 8 | ||||
Oval shape | 6 | 14 | 10 | ||||
Partial confluent necrosis | 55% (11/20) | 15% (3/20) | 10% (2/20) | ||||
Clover shape | 10 | 2 | 1 | |
|||
Maple leaf shape | 1 | 1 | 1 | ||||
Separated necrosis | 15% (3/20) | 0% (0/20) | 0% (0/20) | ||||
Quantitative analysis of coagulation necrosis | |||||||
Circularity (isometric ratio) | 0.69 ± 0.12 | 0.82 ± 0.08 | 0.85 ± 0.08 | <0.001 | <0.001 | <0.001 | 0.758 |
Dmin : Dmax ratio | 0.67 ± 0.19 | 0.79 ± 0.15 | 0.86 ± 0.13 | 0.002 | 0.057 | 0.002 | 0.643 |
Dmax, maximum diameter of the ablation zone; Dmin, minimum diameter of the ablation zone; Dv, vertical diameter of the ablation zone; RFA, radiofrequency ablation.
Values are mean ± standard deviation.
p-values indicate the statistical difference among Groups A–C.
χ2 test results indicating the difference between the three groups.