Table 2.

Summary of the dosimetric results obtained at the audited Centres A and B

Beam dosimetry Measured values at standard distance
Single/dual field at treatment distance
Clinical simulation thermoluminescent dosemeter results
Centre Output (cGy MU−1) Energy ratio Flatness (% average TGAB) Symmetry (%)
Output (cGy MU−1) Energy ratio Mean trunk dose (percentage of prescribed dose ±1 standard deviation)
TG AB
Centre A 10.98 0.650 98.1 99.7 100.8 0.500 (S); 0.465 (D) 0.495 (S); 0.490 (D) 98.4 ± 6.9
(95-cm SSD)
Centre B 12.69 0.529 85.0 101 99.5 0.406 (S); 0.446 (D) 0.511 (S); (D) not measured 93.3 ± 3.6
(100-cm SSD)
Auditing centre baseline values 10.0 0.405 98.5 100 100 0.600 (S and D) 0.465 99.3 ± 8.1 (for all patients to date)
(95-cm SSD)

A, left as facing linac; B, right as facing linac; D, dual; G, gun; MU, monitor unit; S, single; SSD, source to surface distance; T, target.