Table 2.
Beam dosimetry | Measured values at standard distance |
Single/dual field at treatment distance |
Clinical simulation thermoluminescent dosemeter results | |||||
---|---|---|---|---|---|---|---|---|
Centre | Output (cGy MU−1) | Energy ratio | Flatness (% average TGAB) | Symmetry (%) |
Output (cGy MU−1) | Energy ratio | Mean trunk dose (percentage of prescribed dose ±1 standard deviation) | |
TG | AB | |||||||
Centre A | 10.98 | 0.650 | 98.1 | 99.7 | 100.8 | 0.500 (S); 0.465 (D) | 0.495 (S); 0.490 (D) | 98.4 ± 6.9 |
(95-cm SSD) | ||||||||
Centre B | 12.69 | 0.529 | 85.0 | 101 | 99.5 | 0.406 (S); 0.446 (D) | 0.511 (S); (D) not measured | 93.3 ± 3.6 |
(100-cm SSD) | ||||||||
Auditing centre baseline values | 10.0 | 0.405 | 98.5 | 100 | 100 | 0.600 (S and D) | 0.465 | 99.3 ± 8.1 (for all patients to date) |
(95-cm SSD) |
A, left as facing linac; B, right as facing linac; D, dual; G, gun; MU, monitor unit; S, single; SSD, source to surface distance; T, target.