Skip to main content
. 2015 Mar 25;88(1049):20140831. doi: 10.1259/bjr.20140831

Table 3.

Intraobserver and short-term interstudy variability using cardiac MR in normal healthy adults

Parameter Variability Mean difference p-value Limits of agreement Intraclass correlation coefficient (95% confidence interval) Smallest detectable change
LV ejection fraction (%) Intraobserver 1.00 ± 1.27 0.11 −1.48 to 3.48 0.99 (0.94–1.00)
Interstudy 0.57 ± 2.94 0.63 −5.19 to 6.33 0.93 (0.62–1.00) 5.8
LV end-diastolic volume (ml) Intraobserver 0.67 ± 3.62 0.67 −6.43 to 7.77 0.99 (0.95–1.00)
Interstudy −1.57 ± 6.47 0.43 −14.25 to 11.11 0.99 (0.97–1.00) 12.7
LV end-systolic volume (ml) Intraobserver −0.83 ± 2.79 0.50 −6.30 to 4.64 0.99 (0.93–1.00)
Interstudy −1.29 ± 3.65 0.37 −8.44 to 5.86 0.98 (0.91–1.00) 7.2
LV mass (g) Intraobserver −0.33 ± 2.16 0.21 −4.56 to 3.90 0.99 (0.95–1.00)
Interstudy −1.00 ± 3.03 0.50 −6.94 to 4.94 0.98 (0.91–1.00) 5.9

LV, left ventricular.

Data are mean ± standard deviation. Changes in LV parameters were compared using a paired Student's t-test.