. 2015 Apr 3;88(1049):20150034. doi: 10.1259/bjr.20150034

Table 1.

Mean apparent diffusion coefficient (ADC) (± standard deviation) and coefficient of repeatability values obtained in a doped water phantom for each of the diffusion-weighted imaging sequences investigated

Measurement Echoplanar imaging RESOLVE ZOOMit
ADC 1.910 ± 0.026 1.875 ± 0.025 1.891 ± 0.039
Short-term repeatability 0.0155 (0.80%) 0.0125 (0.66%) 0.0597 (3.11%)
Long-term repeatability 0.0620 (3.26%) 0.0667 (3.57%) 0.1183 (6.27%)

All units are ×10−3 mm2 s−1; percentage values are also given with respect to the corresponding mean values for data used in repeatability calculations.