Table 1.

Linearity and sensitivity assessment for DART–MS/MS analysis of CT and NT in human plasma.

Analyte Linearity Correlation coefficient (R2) LOD (μg/mL) LLOQ (μg/mL)
CT Y=0.0923X+1.7316 (0.5–100 μg/mL) 0.9958 0.1 0.3
NT Y=0.1053X+0.0384 (4–100 μg/mL) 0.9994 0.2 0.6