Table 1.
Laser power (W) | ΔT gradient at 25°C (°C) | ΔT gradient at 37°C (°C) |
---|---|---|
3.09×10−1 | −0.668 ln(d/d0)+4.57 | −0.604 ln(d/d0)+4.13 |
7.41×10−1 | −1.76 ln(d/d0)+11.6 | −1.48 ln(d/d0)+9.84 |
1.20 | −2.91 ln(d/d0)+19.2 | −2.23 ln(d/d0)+15.2 |
1.66 | −3.93 ln(d/d0)+26.0 | −2.83 ln(d/d0)+19.6 |
2.14 | −4.79 ln(d/d0)+32.1 | −3.38 ln(d/d0)+23.6 |
Laser power values were those displayed on the controller. Square ROIs (n=51) were set as 5.38 μm×5.38 μm. ΔT gradients were measured three times at 25°C or twice at 37°C. d (in μm) is defined as the distance from the IR laser focal point to the center of the ROI. The unit in the logarithm was cancelled by setting d0=1 (in μm).