Published in final edited form as: Magn Reson Med. 2015 May 2;75(3):1312–1323. doi: 10.1002/mrm.25754

Table 1.

Details of scanning systems used in the study

Log# Vendor Model Field L/d (m/m) Seq.Var. RLb SIb AP(at RL= 0/120mm)c TE (ms) RL/SI
1 Philips Ingenia 3T 1.64/0.7 SSE AX,SAG AX 5/5 115
2 Siemens Verio 3T 1.73/0.7 DSE AX,SAG AX −2/−20 134
3 GEa HDxt-W 1.5T 1.72/0.6 SSE AX,SAG AX 35/37 110
4 GEa HDxt-Z 1.5T 1.72/0.6 SSE AX,SAG AX 7/3 75
5 Siemens Trio 3T 2.13/0.6 DSE AX,SAG AX −5/−25 156
6 Siemens Trio 3T 2.13/0.6 DSE AX AX(p5,6)d −5/−30 103
7 Siemens Espree 1.5T 1.25/0.7 DSE AX,SAG AX −5/−20 135
8 Philips Achieva 3T 1.6/0.6 SSE AX AX(p6)d 15/12 130
9 GE Explorer 3T 1.7/0.6 SSE AX AX 13/5 99
10 Philips Ingenia 3T 1.64/0.7 SSE AX AX 10/10 116/125
11 GE Excite 1.5T 1.72/0.6 S/DSE AX,SAG AX 10/5 96/110
12 Siemens Skyra 3T 1.73/0.7 SSE AX AX −15/−30 158
13 Siemens Espree 1.5T 1.25/0.7 DSE AX AX −35/−57 169
14 Siemens Skyra 3T 1.73/0.7 DSE AX AX −15/−33 175/161

“SSE”: single spin-echo; “DSE”: double spin-echo; “AX”: axial; “SAG”: sagittal; “L/d”: gantry bore length (m) / bore diameter (m)

a

same system with two gradient settings (labeled “W”hole and “Z”oom).

b

image acquisition plane (AX/SAG) for specific phantom alignment direction in the magnet (RL/SI);

c

apparent AP elevation of the phantom at RL offset 0 mm and 120 mm;

d

missing phantom SI-scans (“positions”, see Sup. Mat., Table S1);