Table 3.
CDC DENV1 RT-PCR | |||
negative | Positive | ||
DENV1 RT-PCR | negative | 36 | 0 |
positive | 6 | 43 | |
CDC DENV2 RT-PCR | |||
negative | positive | ||
DENV2 RT-PCR | negative | 67 | 0 |
positive | 6 | 12 | |
CDC DENV3 RT-PCR | |||
negative | positive | ||
DENV3 RT-PCR | negative | 76 | 0 |
positive | 2 | 7 | |
CDC DENV4 RT-PCR | |||
negative | positive | ||
DENV4 RT-PCR | negative | 78 | 0 |
positive | 1 | 6 |
A sample was classified as positive in the CDC DENV-1-4 real-time RT-PCR assays if the amplification curve crossed the threshold line within 37 cycles
Five of the six DENV1 RT-PCR positive samples, two of the six DENV2 RT-PCR positive samples, both the DENV3 RT-PCR positive samples, and the DENV4 RT-PCR positive sample that tested negative in respective CDC DENV-1-4 RT-PCR singleplex assay crossed the threshold line between 37 and 40 cycles in the CDC assays