Table 3.

Monitoring and alarms pertaining to VAVD.

Pressure monitoring
    On hard-shell reservoir 81.7%
    In venous line 5.6%
    None 12.7%
    Stand-alone pressure monitor 22.7%
    Integrated pump console transducer 51.5%
        (without servoregulation)
    Integrated pump console transducer 25.8%
        (with servoregulation)
Alarms and alarm limits
    Positive pressure only 27.4%
    Excessive negative pressure only 6.5%
    Both positive and excessive negative pressure 51.6%
    None 14.5%
    Positive alarm limit (mmHg) +1 to +50
    (mean = +7.71)
    Negative alarm limit (mmHg) −40 to −100
    (mean = −60.83)
Institutional protocol limiting negative pressure
    Yes 61.2%
    No 38.8%
    Institutional protocol limit (mmHg) −40 to −100
    (mean = −52.79)

VAVD, vacuum-assisted venous drainage.