Institutional checklist items pertaining to VAVD.

Item Compliance (%)
Vacuum regulator operational* 53.7
Cardiotomy positive-pressure relief valve present* 55.1
Negative pressure relief valve(s) unobstructed* 33.8
System has been engaged and tested 54.4
Pressure transducers zeroed 81.2
Reservoir pressure reading correlates with vacuum regulator’s set vacuum 26.5
Alarms are on and limits are set properly 71.0

VAVD, vacuum-assisted venous drainage.

*

Checklist item recommended by AmSECT (14)