Table 5.
Item | Compliance (%) |
---|---|
Vacuum regulator operational* | 53.7 |
Cardiotomy positive-pressure relief valve present* | 55.1 |
Negative pressure relief valve(s) unobstructed* | 33.8 |
System has been engaged and tested | 54.4 |
Pressure transducers zeroed | 81.2 |
Reservoir pressure reading correlates with vacuum regulator’s set vacuum | 26.5 |
Alarms are on and limits are set properly | 71.0 |
VAVD, vacuum-assisted venous drainage.
Checklist item recommended by AmSECT (14)