Table 6.
Incident | Prevalence (%) |
---|---|
Pressurized venous reservoir activating PPRV (b) | 23.8 |
Pressurized venous reservoir leading to retrograde air up the venous line | 26.2 |
Air pulled across the membrane during priming | 21.4 |
Leak in vacuum system rendering it non-functional when trying to engage it on CPB | 52.4 |
Vacuum regulator malfunction | 33.3 |
Vacuum source (wall vacuum) malfunction | 33.3 |
Over-administration of drugs or fluids due to increased siphon | 19.1 |