Table 1.

The main features of the five automated immunoassay methods, as declared by the manufacturers

System Method Tracer Antigen Imprecision (%): intra-assay; between-assay Imprecision (%): total LoD (IU/mL) LoQ (IU/mL)
IMM CLIA Luminescence: adamantyl dioxetane phosphate Native TPO 4.9–6.3; nr 6.1–8.1 5.0 nr
MAG CLIA Luminescence: N-(aminobutyl)-N-(ethyl)-isoluminol Native TPO 6.8–7.3; 7.4–7.6 nr 0.38 nr
KRY FIA Fluorescence: europium cryptate/XL 665 Native TPO 2.1–7.6; 5.8–16.0 nr 1.8 9.0
PHA FEIA Fluorescence: β-D-galactosidase; 4-methyl-umbelliferyl-β-D-galactoside Recombinant TPO 4.1–4.7; 2.5–4.8 nr nr nr
LIA CLIA Luminescence: isoluminol derivative Recombinant TPO 3.6–6.2; 4.7–6.9 nr 0.6 1.0

CLIA chemiluminescence immunoassay, FEIA fluorescence enzyme immunoassay, FIA fluoroimmunoassay, IMM Immulite 2000 XPi, KRY Kryptor Compact Plus, LIA Liaison XL, LoD limit of detection, LoQ limit of quantitation, MAG Maglumi 2000 Plus, nr not reported, PHA Phadia 250, TPO thyroid peroxidase