. 2015 Oct;14(5):583–591. doi: 10.7785/tcrtexpress.2013.600280

Table I.

Treatment and dose-volume parameters, monitor units, gantry period, and beam-on times of the four helical tomotherapy plans for patients with stage I lung cancer.

Mean ± standard deviation
2.5-cm Dynamic 2.5-cm Fixed p* 5.0-cm Dynamic 5.0-cm Fixed p** p***
Patient number 10
Total PTV (ml) 22.7 ± 10.3
Pitch 0.215, 0.287
Modulation factor 2.0
Conformity index 1.28 ± 0.12 1.30 ± 0.06 0.32 1.29 ± 0.05 0.16 0.39
Uniformity index 1.09 ± 0.02 1.10 ± 0.02 0.02|| 1.10 ± 0.02 1.11 ± 0.03 0.05|| 0.0002§
Monitor unit 5525 ± 644 5066 ± 635 0.002 3988 ± 245 3219 ± 363 0.002 <0.0001
Gantry period (sec) 38.1 ± 5.8 35.2 ± 5.1 <0.0001|| 41 ± 8.3 32.1 ± 3.9 0.002|| <0.0001§
Beam-on time (sec) 397.9 ± 45.1 365.9 ± 44.6 0.002 290.3 ± 17.4 236.2 ± 25.4 0.002 <0.0001
Body
 V5 Gy (ml) 1825.6 ± 409.4 2143.4 ± 441.6 <0.0001|| 1983.4 ± 401.6 2938.9 ± 559.9 <0.0001|| <0.0001§
 V10 Gy (ml) 806.1 ± 255.2 909.0 ± 271.8 <0.0001|| 881.5 ± 264.7 1196.4 ± 335.7 <0.0001|| <0.0001§
 V20 Gy (ml) 240.3 ± 89.5 269.5 ± 91.3 <0.0001|| 265.6 ± 96.3 354.0 ± 113.2 <0.0001|| < 0.0001§
 V30 Gy (ml) 115.4 ± 43.3 128.0 ± 43.4 <0.0001|| 127.9 ± 46.4 163.4 ± 51.9 <0.0001|| < 0.0001§
Spinal cord maximum (Gy) 10.3 ± 4.6 10.9 ± 4.8 0.003|| 9.2 ± 4.6 8.9 ± 4.9 0.80|| 0.02§
Esophagus maximum (Gy) 14.1 ± 4.1 14.3 ± 4.2 0.14|| 14.0 ± 3.8 14.1 ± 4.1 0.68|| 0.31§

Abbreviation: PTV: Planning target volume.

*p-value between 2.5-cm dynamic- and 2.5-cm fixed- jaw plans.

**p-value between 5.0-cm dynamic- and 5.0-cm fixed- jaw plans.

***p-value among the four plans.

Calculated by Friedman analysis of variance.

Calculated by Wilcoxon signed-rank test.

§Calculated by parametric analysis of variance for depended samples.

||Calculated by paired t-test.