Skip to main content
. 2015 May 28;4(Suppl 1):51–58. doi: 10.1159/000430840

Table 3.

Intraday validation outcome for TH and TAM extraction

Intraday (n = 10)/ analytes HQAL (80 nM)
MQAL (10 nM)
LQAL (0.75 nM)
mean conc., nM CV, % accuracy, % mean conc., nM CV, % accuracy, % mean conc., nM CV, % accuracy, %
T0 82.61 3.7 103.3 12.62 2.1 126.2 2.93 1.9 391.1
3-T1 77.08 2.7 96.4 10.62 1.9 106.2 0.82 4.1 108.7
3′-T1 73.07 2.6 91.3 10.23 2.9 102.3 0.82 3.9 109.7
3,5-T2 71.71 1.6 89.6 10.16 3.1 101.6 0.80 2.7 106.6
3,3′-T2 73.67 2.7 92.1 9.94 2.9 99.4 0.79 3.6 104.7
3′,5′-T2 75.24 2.3 94.0 10.51 3.3 105.1 0.81 3.9 107.8
T3 71.21 2.4 89.0 9.70 3.0 97.0 0.76 3.1 100.7
rT3 78.45 2.3 98.1 10.36 2.6 103.6 0.77 3.1 102.4
T4 72.20 2.0 90.2 9.62 4.1 96.2 0.72 6.4 96.5

t4am 90.07 4.6 112.6 11.60 2.5 116.0 0.89 5.1 119.2
t3am 84.32 1.8 105.4 11.34 2.5 113.4 0.88 3.5 116.7
rT3AM 83.66 2.1 104.6 11.35 2.9 113.5 0.86 4.5 114.0
3,5-T2AM 83.55 2.0 104.4 11.25 2.2 112.5 0.86 3.5 115.2
3-T1AM 82.72 2.6 103.4 11.31 2.8 113.1 0.88 3.1 117.3
T0AM 88.76 1.2 111.0 11.50 1.5 115.0 1.02 3.9 111.2

R2 = Coefficient of determination; mean conc. = mean determined concentration. DMEM/F12 contains approximately 1.54 nM T0 as endogenous background concentration.