Skip to main content
. 2015 Nov 12;4(4):323–342. doi: 10.5501/wjv.v4.i4.323

Table 3.

Performance characteristics of commercial methods for hepatitis B virus and hepatitis C virus detection and quantification

Assay (manufacture) Method Measurable range (IU/mL) Limit of detection IU/mL (using WHO HBV standard) Conversion factor (IU/mL to copies/mL)
HBV
Cobas ampliPrep/Cobas TaqMan HBV test v2.0 (Roche Diagnostics, California, United States) Semi-automated qPCR 20-1.7 × 107 20 5.82
Cobas TaqMan HBV test (for use with high pure system) Semi-automated real time PCR 29-1.1 × 107 6 5.82
Abbott real time HBV (Abbott Diagnostic, Chicago, United States) Automated real time PCR 10-1 × 109 10 3.41
Versant HBV 3.0 assay (Siemens Healthcare, United States) Branched DNA 2000-1 × 108 2000 5.6
HCV qualitative
Cobas Amplicor HCV v2.0 (Roche) Semiautomated RT-PCR 50 50 -
Versant HCV RNA qualitative assay (TMA) Transcription-mediated amplification < 9.6 5-9.6 -
Quantitative
Versant HCV RNA 3.0 (Siemens) Branched DNA 7.7 × 106 615 5.2
Cobas Amplicor Monitor HCV v2.0 (Roche) Semiautomated RT-PCR 5 × 105 600 2.7
Real time HCV (Abbott) Real time PCR 1 × 107 10 3.8
Cobas AmpliPrep/Cobas TaqMan (Roche) Automated real time PCR 6.9 × 107 43 3
High Pure/Cobas TaqMan (Roche) Semiautomated real time PCR 3.9 × 108 25 3

WHO: World Health Organization; HBV: Hepatitis B virus; HCV: Hepatitis C virus; PCR: Polymerase chain reaction; RT-PCR: Reverse transcriptase-PCR; qPCR: Quantitative PCR.