Table 1.

Factory installed protocols on the CS 9300 CBCT machine. Each patient setting was preprogrammed by the manufacturer, varying the kVp, mA, and reconstructed voxel size for each patient size. These are the settings that are used clinically.

FOV cm2 (diameter × height) Patient type kVp mA Time (s) Voxel (μm)
FOV 17×11 Child 80 4 6.40 250
Adult small 85 4 6.40 250
Adult average 90 4 6.40 250
Adult large 90 5 6.40 250
FOV 17×6 Child 80 4 6.20 200
Adult small 85 4 6.30 200
Adult average 90 4 6.30 200
Adult large 90 5 6.30 200
FOV 8×8 Child 80 4 8.00 180
Adult small 85 4 8.00 180
Adult average 90 4 8.00 180
Adult large 90 5 8.00 180
FOV 10×10 Child 80 4 6.20 180
Adult small 85 4 6.20 180
Adult average 90 4 6.20 180
Adult large 90 5 6.20 180
FOV 10×5 Child 80 4 6.20 180
Adult small 85 4 6.20 180
Adult average 90 4 6.20 180
Adult large 90 5 6.20 180
FOV 5×5 Child 75 4 12 200
Adult small 80 5 12 200
Adult average 84 5 12 200
Adult large 84 8 12 200