. 2015 Nov 18;2(4):044002. doi: 10.1117/1.JMI.2.4.044002

Table 2.

Noise, uniformity, geometric distortion, and AAV obtained for different protocols on CS 9300 CBCT.

FOV cm2 (diameter×height) Patient type Noise (%) Uniformity (%) Geometric distortion (mm) AAV
17×11 Child 19.5 93.5 9.98 0.4
Adult small 17.9 93.9 9.97 0.4
Adult average 16.4 94.8 9.99 0.3
Adult large 13.7 96.9 10.01 0.3
17×6 Child 9.9 98.5 9.97 0.3
Adult small 9.1 99.1 9.99 0.3
Adult average 8.2 99.7 9.98 0.2
Adult large 7.5 98.8 9.98 0.3
8×8 Child 7.6 94.4 10.00 0.3
Adult small 7.3 95.6 10.00 0.3
Adult average 6.1 98.3 10.00 0.3
Adult large 6.4 98.4 10.05 0.3
10×10 Child 9.5 93.6 10.01 0.3
Adult small 8.8 94.2 10.02 0.3
Adult average 8.2 94.3 10.04 0.3
Adult large 7.1 96.0 9.95 0.3
10×5 Child 8.6 95.8 10.02 0.4
Adult small 8.6 95.9 10.02 0.4
Adult average 7.9 96.1 10.02 0.4
Adult large 7.1 98.5 10.01 0.3
5×5 Child 12.5 97.5 10 0.4
Adult small 10.8 96.6 10 0.4
Adult average 9.4 96.6 10 0.4
Adult large 8.5 95.3 10 0.4