Table 2 Sensitivity and specificity of all three systems (system 1: Inspiration™; system 2: BioVision™, and system 3: Tomosynthesis) as determined by 3 radiologists through the prospective assessment of 600 specimen radiograms/projection slides (cine mode) from 200 patients with a total of 200 malignant lesions.

Sensitivity (%) Specificity (%)
Magnification1.0 : 1.0 (all 3 systems) System 1 Inspiration (85 µm) System 2 BioVision (50 µm) System 3 Tomosynthesis (85 µm) System 1 Inspiration (85 µm) System 2 BioVision (50 µm) System 3 Tomosynthesis (85 µm)
Radiologist 1 78.0 79.0 86.0 100.0 100.0 100.0
Radiologist 2 79.0 78.0 87.0 100.0 100.0 100.0
Radiologist 3 78.0 78.0 87.0 100.0 100.0 100.0
Mean 78.3 78.3 86.6 100.0 100.0 100.0