Table 4.

Measured artefact volume (cm3) and the according percentage of artefact reduction for each evaluated sequence compared with conventional acquisition based on four image contrasts

Magnetic field Metallic implant Product VAT Slice encoding for metal artefact correction-VAT
1.5 T SS screw 29.3 ± 18.8 9.9 ± 6.4 (65% ± 10%) 3.8 ± 0.8 (83% ± 9%)
SS plate 279.6 ± 75.9 80.4 ± 24.6 (71% ± 7%) 30.8 ± 9.7 (89% ± 3%)
3.0 T SS screw 57.5 ± 23.2 30.0 ± 10.1 (45% ± 15%) 17.0 ± 10.9 (72% ± 7%)
SS plate 480.4 ± 49.2 394.4 ± 76.9 (19% ± 11%) 304.4 ± 87.9 (38% ± 13%)

Product, conventional sequence; SS, stainless steel; VAT, view angle tilting.

The results are expressed as mean ± standard deviation.