Table 2.

Comparison of two strategies by various parameters

Variable Non-coplanar intensity-modulated radiotherapy plan Stereotactic body radiotherapy plan p-value
Clinical target volume (ml) 9.99 (range, 1.16–42.82)
 
PTV volume (ml) 34.84 (range, 6.59–88.85)
 
PTV coverage (%)
V100% 86.9 ± 12.0 86.7 ± 15.5 0.59
V95% 98.9 ± 2.8 99.3 ± 0.9 0.20
V90% 99.7 ± 0.9 99.9 ± 0.2 0.34
Radiation Therapy Oncology Group CI 1.05 ± 0.21 1.47 ± 0.21 <0.001
Healthy tissue CI 0.84 ± 0.09 0.63 ± 0.14 <0.001
CN
 CN 100 0.72 ± 0.10 0.53 ± 0.13 <0.001
 CN 75 0.33 ± 0.08 0.27 ± 0.06 <0.001
 CN 50 0.18 ± 0.04 0.15 ± 0.04 <0.001
Homogeneity index 1.04 ± 0.03 1.11 ± 0.03 <0.001

CI, conformity index; CN, conformal number; PTV, planning target volume; Vx%, the volume receiving ≥x% of the prescribed dose.

Parameters are calculated by Wilcoxon signed-rank sum test.