Table 1.

CT scan parameters used in the study, for the four different CT scanners

Parameter Philips Ingenuity Core (single energy) [Philips (Cleveland, OH)] Toshiba Aquilion ONE™ Vision Edition (single energy) [Toshiba (Otawara, Japan)] GE Discovery™ 750HD (dual energy by fast kilovoltage switching) [GE (Milwaukee, WI)] Siemens SOMATOM® Definition Flash (dual energy with dual sources) [Siemens (Forchheim, Germany)]
CT protocol Helical Volume Helical Helical
Collimation (mm) 64 × 0.625 280 × 0.5 64 × 0.625 64 × 0.6
MAR technique MAR algorithm (O-MAR) MAR algorithm (SEMAR) Monoenergetic reconstruction (110 keV)
MAR algorithm (MARS)
Monoenergetic reconstruction (110 keV)
DE-composition reconstructions (weight of −0.3)
Iterative reconstruction iDose AIDR 3D (adaptive iterative dose reduction 3D) ASIR (adaptive statistical iterative reconstruction) SAFIRE (sinogram affirmed iterative reconstruction)
Level 3 (range, 1–5) Level standard (range, mid, standard and strong) Level 50% (range, 0–100%) Level 3 (range, 1–5)
Soft kernel B FC08 Standard D34 (FB)
Q30 (IR)
Sharper kernel YB FC30 Detail D45 (FB)
Q50 (IR)

FB, filtered backprojection; IR, iterative reconstruction; MAR, metal artefact reduction; MARS, metal artefact reduction software; O-MAR, MAR for orthopaedic implants; SEMAR, single-energy MAR.