Skip to main content
. 2015 Nov 18;53(12):3822–3827. doi: 10.1128/JCM.02116-15

TABLE 3.

Differences between BKV DNA loads measured by the three methods according to sample type and viral genotype

Method comparison Mean (SD) difference between BKV DNA loads (log10 copies/ml)
All samples Plasma samples Urinary samples Genotype I Genotype IV
R-gene vs GeneProof −0.08 (0.46) −0.17 (0.25) −0.03 (0.54) −0.05 (0.46) −0.12 (0.44)
R-gene vs RealStar −0.27 (0.48) −0.55 (0.36)a −0.12 (0.47) −0.25 (0.53) −0.33 (0.38)
GeneProof vs RealStar −0.19 (0.51) −0.38 (0.34) −0.09 (0.56) −0.20 (0.52) −0.22 (0.49)
a

P < 0.05.