. 2015 Nov 18;53(12):3822–3827. doi: 10.1128/JCM.02116-15

TABLE 4.

Quantification of BKV DNA loads of AcroMetrix BKV panel specimens

AcroMetrix BKV panel specimen no.a VL result (log10 copies/ml)
Difference from expected VL result (log10 copies/ml)
Expected R-gene GeneProof RealStar R-gene GeneProof RealStar
1 2.70 2.30 2.50 2.63 −0.40 −0.20 −0.07
2 3.70 3.92 3.98 3.90 0.22 0.28 0.20
3 4.70 4.74 4.76 4.70 0.04 0.06 0.00
4 5.70 5.54 5.77 5.56 −0.16 0.07 −0.14
5 6.70 6.74 7.06 6.70 0.04 0.36 0.00
Mean difference −0.052 0.112 −0.002
a

Three samples of each specimen were tested.