Skip to main content
. 2008 Oct 1;113(5):265–280. doi: 10.6028/jres.113.020

Table 1.

Calculated 68Ga activity from various γ-ray gates and using Eq. (3), with and without the final correction term, and relative to the corrected G2 value, (used for the final activity determination). The uncertainties (k = 1) on the uncorrected activities are standard deviations of the intercepts from the least-squares fits to the various data sets. The uncertainties (k = 1) of the corrected values are from estimates of the uncertainties in εann and εγ1077

γ-ray gate Uncorrected N0 (Bq) Corrected N0 (Bq)
G1 - Comptons 1.020 ± 0.002 1.002 ± 0.004
G2 - photopeak 1.0021 ± 0.001 1.0000 ± 0.0004
G3 - sum peak 1.033 ± 0.002 1.000 ± 0.007